A reduced-order modeling for efficient design study of artificial valve in enlarged ventricular outflow tracts - Inria - Institut national de recherche en sciences et technologies du numérique Access content directly
Journal Articles Computer Methods in Biomechanics and Biomedical Engineering Year : 2016

A reduced-order modeling for efficient design study of artificial valve in enlarged ventricular outflow tracts

Abstract

A computational approach is proposed for efficient design study of a reducer stent to be percutaneously implanted in enlarged right ventricular outflow tracts (RVOT). The need for such a device is driven by the absence of bovine or artificial valves which could be implanted in these RVOT to replace the absent or incompetent native valve, as is often the case over time after Tetralogy of Fallot repair. Hemodynamics are simulated in the stented RVOT via a reduce order model based on proper orthogonal decomposition, while the artificial valve is modeled as a thin resistive surface. The reduced order model is obtained from the numerical solution on a reference device configuration, then varying the geometrical parameters (diameter) for design purposes. To validate the approach, forces exerted on the valve and on the reducer are monitored, varying with geometrical parameters, and compared with the results of full CFD simulations. Such an approach could also be useful for uncertainty quantification.
Fichier principal
Vignette du fichier
short-pod-2-HAL.pdf (1.53 Mo) Télécharger le fichier
Origin : Files produced by the author(s)
Loading...

Dates and versions

hal-01414386 , version 1 (21-12-2016)

Identifiers

Cite

Alfonso Caiazzo, Romain Guibert, Irene E. Vignon-Clementel. A reduced-order modeling for efficient design study of artificial valve in enlarged ventricular outflow tracts. Computer Methods in Biomechanics and Biomedical Engineering, 2016, 19 (12), pp.1314 - 1318. ⟨10.1080/10255842.2015.1133811⟩. ⟨hal-01414386⟩
501 View
220 Download

Altmetric

Share

Gmail Facebook X LinkedIn More