Transit time ultrasound perivascular flow probe technology is superior to MR imaging on hepatic blood flow measurement in a porcine model - Inria - Institut national de recherche en sciences et technologies du numérique
Article Dans Une Revue Hepatobiliary and Pancreatic Diseases International Année : 2018

Transit time ultrasound perivascular flow probe technology is superior to MR imaging on hepatic blood flow measurement in a porcine model

Résumé

BACKGROUND: The hepatic hemodynamics is an essential parameter in surgical planning as well as in various disease processes. The transit time ultrasound (TTUS) perivascular flow probe technology is widely used in clinical practice to evaluate the hepatic inflow, yet invasive. The phase-contrast-MRI (PC-MRI) is not invasive and potentially applicable in assessing the hepatic blood flow. In the present study, we compared the hepatic inflow rates using the PC-MRI and the TTUS probe, and evaluated their predictive value of post-hepatectomy adverse events. METHODS: Eighteen large white pigs were anaesthetized for PC-MRI and approximately 75% hepatic resection was performed under a unified protocol. The blood flow was measured in the hepatic artery (Qha), the portal vein (Qpv), and the aorta above the celiac trunk (Qca) using PC-MRI, and was compared to the TTUS probe. The Bland-Altman method was conducted and a partial least squares regression (PLS) model was implemented. RESULTS: The mean Qpv measured in PC-MRI was 0.55 ± 0.12 L/min, and in the TTUS probe was 0.74 ± 0.17 L/min. Qca was 1.40 ± 0.47 L/min in the PC-MRI and 2.00 ± 0.60 L/min in the TTUS probe. Qha was 0.17 ± 0.10 L/min in the PC-MRI, and 0.13 ± 0.06 L/min in the TTUS probe. The Bland-Altman method revealed that the estimated bias of Qca in the PC-MRI was 32% (95% CI: -49% to 15%); Qha 17% (95% CI: -15% to 51%); and Qpv 40% (95% CI: -62% to 18%). The TTUS probe had a higher weight in predicting adverse outcomes after 75% resection compared to the PC-MRI (β= 0.35 and 0.43 vs β = 0.22 and 0.07, for tissue changes and premature death, respectively). CONCLUSIONS: There is a tendency of the PC-MRI to underestimate the flow measured by the TTUS probes. The TTUS probe measures are more predictive of relevant post-hepatectomy outcomes.
Fichier non déposé

Dates et versions

hal-01954795 , version 1 (14-12-2018)

Identifiants

Citer

Mohamed Bekheit, Chloe Audebert, Petru Bucur, Hans Adriansen, Emilie Bled, et al.. Transit time ultrasound perivascular flow probe technology is superior to MR imaging on hepatic blood flow measurement in a porcine model. Hepatobiliary and Pancreatic Diseases International, 2018, 17 (6), pp.538-545. ⟨10.1016/j.hbpd.2018.07.009⟩. ⟨hal-01954795⟩
165 Consultations
0 Téléchargements

Altmetric

Partager

More