Off-resonance correction of non-Cartesian SWI using internal field map estimation
Correction d'inhomogénéités de champs pour la SWI non-cartésienne par estimation des cartes de champs
Résumé
Patient-induced inhomogeneities in the magnetic field cause distortions and blurring during acquisitions with long echo times, as in susceptibility-weighted imaging. Most correction methods require collecting an additional ΔB0 field map. To avoid that, we propose a method to approximate this field map using the single echo acquisition only. The main component of the observed phase is linearly related to ΔB0 and TE, and the relative impact of non-ΔB0 terms becomes insignificant with TE>20ms at 3T. The estimated 3D field maps, produced at 0.6 mm isotropic under 3 minutes, provide equivalent corrections to acquired ones.
Les inhomogénéités de champs induites par les patients sont à l'origine de distorsions et de floutages durant les acquisitions à temps d'écho longs, comme pour l'imagerie pondérée en susceptibilité. La plupart des méthodes de correction nécessitent d'acquérir une carte de champ ΔB0 additionnelle. Pour éviter cela, nous proposons une méthode pour approximer cette carte de champs en utilisant seulement l'acquisition à écho unique. La composante principale de la phase observée est linéairement liée au ΔB0 et au TE, et l'impact relatif des termes indépendants du ΔB0 deviennent négligeables pour TE>20ms à 3T. Les cartes 3D estimées, produites à 0.6 mm isotrope en moins de 3 minutes, permettent d'obtenir une correction équivalente aux cartes acquises.